applanation tonometer
Accurate Measurement
With years of tests and clinical applicationsGoldmann has been proved to be the the gold standard for tonometry measurement. Based on the proven evidenceMediWorks T170 tonometer applied Goldmann principle in the design and manufacture.
With precision manufacturing process,MediWorks applanation tonometer offers accurate and consistent measurements which give a clear picture of patientslOP situation to the doctors
Excellent Mechanics
Full attentions are paid to every detail in the mangfacture of
tonometers.Workers assemble every
component carefully and check tight tolerances in weight to make sure the tonometer deliver consistent measurements
The product is designed on the principle basis of Goldman tonometer.Three types are available:R and T Type are suitable for Haig Streit type slit lamp,while Z type is suitable for Carl Zeiss type slit lamp.
ophthalmic tonometer
Characteristics:
It's not only used to examine the eyes, but also measure the intraocular pressure while working under the slit lamp.
Accurate measurement ensures the total tolerrance less than ± 0.066KPa (0.5mmHg).
Directly get the intraocular pressure.
Measured intraocular pressure will not be affected by hadrness of eye.
Adjustable measuring pressure ensures the working stability and accurate measurement of tonometer in long time.
Mrasurement Range: 0-10.64 KPa
Light Ring Displacement: 1.53*2=3.06mm
Diameter of the Testing head: 7mm
Average Error: +/-70 Pa